High concurrent validity between digital and analogue algometers to measure pressure pain thresholds in healthy participants and people with migraine: a cross-sectional study

Abstract

Background

Pressure pain thresholds (PPTs) are commonly assessed to quantify mechanical sensitivity in various conditions, including migraine. Digital and analogue algometers are used, but the concurrent validity between these algometers is unknown. Therefore, we assessed the concurrent validity between a digital and analogue algometer to determine PPTs in healthy participants and people with migraine.

Methods

Twenty-six healthy participants and twenty-nine people with migraine participated in the study. PPTs were measured interictally and bilaterally at the cephalic region (temporal muscle, C1 paraspinal muscles, and trapezius muscle) and extra-cephalic region (extensor carpi radialis muscle and tibialis anterior muscle). PPTs were first determined with a digital algometer, followed by an analogue algometer. Intraclass correlation coefficients (ICC3.1) and limits of agreement were calculated to quantify concurrent validity.

Results

The concurrent validity between algometers in both groups was moderate to excellent (ICC3.1 ranged from 0.82 to 0.99, with 95%CI: 0.65 to 0.99). Although PPTs measured with the analogue algometer were higher at most locations in both groups (p < 0.05), the mean differences between both devices were less than 18.3 kPa. The variation in methods, such as a hand-held switch (digital algometer) versus verbal commands (analogue algometer) to indicate when the threshold was reached, may explain these differences in scores. The limits of agreement varied per location and between healthy participants and people with migraine.

Conclusion

The concurrent validity between the digital and analogue algometer is excellent in healthy participants and moderate in people with migraine. Both types of algometer are well-suited for research and clinical practice but are not exchangeable within a study or patient follow-up.

Background

Pressure pain threshold (PPT) is defined as the amount of pressure applied over the skin when a feeling of pressure transitions into a feeling of painful pressure [1]. PPTs are assessed to determine increased mechanical sensitivity as a feature of central and/or peripheral sensitization in headache research [2, 3]. Significantly lower PPTs are reported in people with migraine compared to healthy participants at the cephalic region [1], but the evidence is inconsistent for the extra-cephalic region [4, 5]. Algometry is not only used in research to elucidate the underlying neurophysiological mechanisms. In clinical practice, PPT measurements are also used to assess mechanical pressure sensitivity in people with headache disorders and evaluate treatment effects [6,7,8,9].

Most studies in headache research [1,2,3,4] used a digital algometer to measure PPTs. The main difference between a digital and analogue algometer is the use of a digital display versus a dial with a needle to determine the PPT value [10]. Some digital algometers (e.g. Type II, Somedic Electronics, Solna, Sweden) has a hand-held switch which the participant can when the threshold is reached. For some analogue algometers (Force Dial FDK, Wagner Instruments, Greenwich, Connecticut) the examiner responds to a verbal cue from the participant. Furthermore, some digital algometers (e.g., Type II, Somedic Electronics, Solna Sweden) provide visual feedback about the pressure application rate. The pressure can be expressed in kilo Pascals (kPa) or kilogram per square centimetre (kg/cm2). Another main difference is the acquisition cost, with an analogue algometer typically being considerably less expensive (e.g., ~$280 for a Wagner analogue algometer and ~ $5.500 for a Somedic digital algometer). The clinimetric properties of analogue and digital algometers that have been reported show excellent construct validity [11], high intrarater [12, 13] and interrater [14, 15] reliability in healthy participants and people with headache [16]. Although both algometers show high construct validity and reliability, we lack information about the concurrent validity between these tools in healthy participants and people with migraine. If researchers and clinicians want to be able to compare the findings in the literature or across clinical practices, they need to know how valid analogue systems are compared to digital systems.

Therefore, this study aimed to estimate the concurrent validity between digital and analogue algometers in healthy participants and people with migraine in cephalic and extra-cephalic regions.

Methods

Design

This study was a prospective observational study on healthy participants and people with migraine. The Medical Ethical Committee of the Amsterdam University Medical Centre approved the study (METC-2015-551). Written informed consent was obtained from all participants prior to the start of the measurements.

Participants

Healthy participants were recruited from the general population via advertising. Healthy participants were excluded if they reported any history of neck pain, headache, chronic pain, psychiatric conditions or neurological disorders in the last 12 months.

In the same period, people with migraine were recruited from primary healthcare clinics in The Netherlands. People with migraine were eligible for inclusion if they were: diagnosed with migraine according to the International Classification of Headache Disorders (ICHD - III) [17], between 18 and 65 years old, and Dutch or English speaking. Exclusion criteria were: participants with other types of headaches, such as medication overuse headache, a non-specific cause of headache, traumatic head/neck symptoms within 2 months before the measurements, chronic musculoskeletal painful conditions, psychiatric conditions, malignancy or neuropathic pain states other than migraine. Participants were not allowed to use analgesic medications or non-steroid anti-inflammatory drugs (NSAID) within 24 h before the measurements or receive treatment within 48 h before the measurements.

Assessor

One assessor (TD) performed all measurements. During a 2 weeks period, the assessor underwent training for ~ 15 h to learn how to calibrate the instruments, maintain the required rate of pressure increase and apply the standardized protocol. The training PPT measurements were performed on healthy adult volunteers who did not participate in the actual experiment.

Questionnaires

At baseline, all participants completed a general questionnaire to document sociodemographic information. They scored their current headache intensity on a 10-point Numeric Pain Rating Scale (NPRS, 0 = no pain and 10 = worst possible pain) and completed the Headache Impact Test-6 (HIT-6) questionnaire. The NPRS has been shown to be a valid and reliable instrument in several populations [18]. The HIT-6 questionnaire was used to measure the influence of headache on daily functioning. Scores on the HIT-6 questionnaire range from 36 to 78: higher scores indicate a higher impact of migraine. The HIT-6 questionnaire is a reliable (ICC = 0.77) and valid tool to determine headache impact in a migraine population [19,20,21].

Concurrent validity

Healthy participants and people with migraine were measured to determine the concurrent validity at both the cephalic and extra-cephalic regions. Since pain sensitivity differs between sides in people with migraine [21], concurrent validity was determined separately for the dominant and non-dominant side of the migraine. The dominant side was defined as the most painful side of the last migraine attack. The side of the dominant hand was used to define the corresponding ‘dominant side’ in healthy participants. People with migraine were measured during their interictal phase (the phase without headache or other symptoms between two consecutive migraine attacks). To ensure that the examination was in the interictal phase, we contacted the participants before the initial measurement. All participants with migraine were asked when they had their last ictal and post-ictal phase and if they were in a headache-free period for more than 48 h.

Digital algometer

A digital algometer (Type II, Somedic Electronics, Solna, Sweden) with a 1 cm2 probe was used to measure PPTs (Fig. 1A). The Somedic algometer has excellent construct validity (ICC > 0.98) compared to a force plate [11]. The ICC values of intrarater reliability are ranging from 0.90 to 0.95 in healthy participants and from 0.89 to 0.96 in people with migraine [22, 23].

Fig. 1
figure 1

A Differences Pressure Pain Thresholds digital and analogue algometer Temporalis muscle dominant side in healthy participants. B Differences Pressure Pain Thresholds digital and analogue algometer Temporalis muscle dominant side in people with migraine

Analogue algometer

A Wagner algometer (Force Dial FDK 20, Wagner Instruments, Greenwich. Connecticut) with a 1 cm2 probe area was used as an analogue algometer (Fig. 1B). The hand-held Wagner algometer with a digital reading function shows comparable correlations to a force plate (Pearson’s r = 0.99) [24]. Previous studies with comparable hand-held analogue algometers showed ICC values for intrarater reliability ranging from 0.79 to 0.91 in healthy participants [14] and from 0.90 to 0.97 for people with chronic tension-type headache [16].

Measurement procedures

Prior to commencing the experiment, the Somedic and Wagner algometers were calibrated according to the manufacturer’s instructions. To familiarise participants with the procedures, two PPT measurements were performed for both the digital and analogue algometer at the midpoint of the biceps brachii muscle. Throughout the experiment, participant sat in a relaxed position on a chair, in an upright position, with both arms supported by armrest. At the start of the measurement, all locations were marked on the participants’ skin with a black skin marker. Each location was measured three times with the analogue and the digital algometer. There was a 30 min interval between the measurements with the digital and analogue algometer, and a 30s interval between the three consecutive repetitions to minimize wind-up. To reduce variability, we placed the tip of both algometers in the same place, i.e., the footprint of the tip of the algometer at the marked location which remained visible between the 30s repetitions.

PPTs were assessed bilaterally at five test locations in a standardized order: from the most cranial location to the most caudal location, i.e.: (1) temporal muscle (1 cm lateral to the external angle of the orbit), (2) C1 paraspinal muscles (2 cm lateral from the midline of the neck), (3) upper trapezius muscle (on the midpoint between the acromion and spinous process C7), (4) extensor carpi radialis muscle (at 1/3 length of the forearm distal to the elbow) and (5) tibialis anterior muscle (at 1/3 of the length of the lower leg, distal to the knee). In people with migraine, all measures were first performed on the dominant side of the migraine.

PPTs were first measured with the Somedic algometer. The assessor placed the tip of the algometer perpendicular on the skin over the target location and increased the pressure with a rate of 50 kPa/s [3, 25]. Participants were instructed to press the switch when the feeling of pressure changed into the feeling of pain. Once the button was pressed, the measurement stopped automatically. Once all measurements from all locations were obtained with the digital algometer, the measurements with the analogue algometer were performed by the same assessor. Measurements with the analogue algometer were performed at the same locations, visible by the ‘pressure mark’ from the tip of the algometer from the previous measurements. The assessor was trained to increase the pressure with the Wagner algometer at a rate of 0.5 kg/cm2/s [3, 25]. Participants had to say ‘stop’ when the threshold was reached as hand-held switches are not available for analogue algometers. After each measurement, the assessor read the pressure from the display of the analogue algometer and entered this score in a datasheet.

Blinding

The assessor was blinded for the participant’s condition. All participants were blinded for the measurement outcomes. The data analysis was performed by a researcher who was not involved in the measurements.

Data analysis

The mean of the three PPT measures was calculated for each location for each participant [26]. Values from the analogue algometer were converted into kPa using the following formula: 1 kg-force per square centimetre (kg/cm2) = 98.0665 kPa.

Statistical analysis

Because no sample size calculation for concurrent validity is available [27], we used a comparable number of participants as used in other concurrent validity studies [28, 29]. Statistical analyses were performed using SPSS version 25.0 (SPSS Inc., Chicago, Illinois, USA). Normality of continuous variables was checked by visual inspection of the Q-Q plots and Shapiro-Wilk test. Mean differences between digital and analogue measurement were analyzed with an independent samples t-test. We used a two-way mixed effect, consistency and single rater/measurement model (ICC3.1) to quantify the concurrent validity between the algometers [30]. The following cut-off points were used for interpretation of the ICC: < 0.5: poor validity, 0.5–0.75: moderate validity, 0.75–0.9: good validity, and values > 0.90: excellent validity [31]. Additionally, limits of agreement (= d ± 1.96 x SD) were calculated with Bland-Altman plots to determine the mean and 95% confidence interval (CI) of the differences between the digital and analogue algometer [32]. The independent samples t-test was performed to test systematic differences in PPTs between the digital and analogue algometer.

Results

Twenty-six healthy participants (69% female; mean (SD) age: 42.8 (14.3) years) and 29 people with migraine (89% female; 47.8 (10.8) years) were included in this study. Table 1 provides an overview of the participants’ characteristics. There were no missing data.

Table 1 Main characteristics at baseline of healthy participants and people with migraine

Table 2 presents the mean PPT values of the digital and analogue algometer for each test locations at the dominant and non-dominant side of healthy participants and people with migraine. Compared to the digital algometer higher PPT values were obtained with the analogue algometer, indicating a systematic error between the digital and analogue PPT measurements. In both groups, the mean difference in PPT values between both algometers ranged from − 6.3 to 2.3 kPa in healthy participants and from − 18.3 to − 3.5 kPa in people with migraine (Table 3).

Table 2 Pressure pain thresholds and concurrent validity in healthy participants and people with migraine
Table 3 Mean difference of all locations between the digital and analogue algometer and limits of agreement in people with healthy participants and people with migraine

Concurrent validity

The concurrent validity in the cephalic and extra-cephalic region in healthy participants was excellent (ICC 3.1 > 0.97). In people with migraine, the ICCs were good to excellent and ranged from 0.82 to 0.97 (Table 2). Limits of agreement (LoA) are shown in Table 3 and displayed in the Bland & Altman plots (Figs. 2A-B, 3A-B, 4A-B, 5A-B, 6A-B).

Fig. 2
figure 2

A Differences Pressure Pain Thresholds digital and analogue algometer Temporalis muscle dominant side in healthy participants (mean difference, upper and lower limit). B Differences Pressure Pain Thresholds digital and analogue algometer Temporalis muscle dominant side in people with migraine (mean difference, upper and lower limit)

Fig. 3
figure 3

A Differences Pressure Pain Thresholds digital and analogue algometer C1 paraspinal muscles dominant side in healthy participants (mean difference, upper and lower limit). B Differences Pressure Pain Thresholds digital and analogue algometer C1 paraspinal muscles dominant side in people with migraine (mean difference, upper and lower limit)

Fig. 4
figure 4

A Differences Pressure Pain Thresholds digital and analogue algometer Trapezius muscle dominant side in healthy participants (mean difference, upper and lower limit). B Differences Pressure Pain Thresholds digital and analogue algometer Trapezius muscle dominant side in people with migraine (mean difference, upper and lower limit)

Fig. 5
figure 5

A Differences Pressure Pain Thresholds digital and analogue algometer Extensor carpi radialis muscle dominant side in healthy participants (mean difference, upper and lower limit). B Differences Pressure Pain Thresholds digital and analogue algometer Extensor carpi radialis muscle dominant side in people with migraine (mean difference, upper and lower limit)

Fig. 6
figure 6

A. Differences Pressure Pain Thresholds digital and analogue algometer Tibialis anterior muscle dominant side in healthy participants (mean difference, upper and lower limit) B. Differences Pressure Pain Thresholds digital and analogue algometer Tibialis anterior muscle dominant side in people with migraine (mean difference, upper and lower limit)

Discussion

The Somedic digital and Wagner analogue algometer showed good to excellent concurrent validity in healthy participants and in people with migraine. At most locations, the PPT scores obtained with the analogue algometer were higher, suggesting a systematic error. Although we found good to excellent concurrent validity, the upper and lower limits of the mean difference between the digital and the analogue algometer reflect a wide range in scores in people with migraine. This means that clinicians have to be aware that PPTs measured with a digital or analogue algometer may differ substantially in people with migraine.

The difference in the method of digital and analogue measurement may have caused the systematic error. When the participant pressed the hand-held switch, the reading of the threshold on the digital algometer stopped immediately. During the test with the analogue algometer, the participant had to give a verbal cue to the assessor. Then, the assessor stopped the test. The time needed for the assessor to stop the test may have resulted in a slightly higher pressure score. This delay in terminating the measurement seems to be the most plausible reason for the systematically higher PPT values retrieved with the analogue algometer. Besides this difference in method of measurement, the fixed order in which we first measured all locations with the digital algometer may have contributed to systematic error. Although randomization of the order of measurement would be preferable, we did not randomize the order of the measurements as this study was part of a larger project evaluating PPTs in people with migraine by using a digital algometer [33]. Randomization of the algometers (analogue versus digital) would have strengthened the design of this study.

Still, randomization of measurement order appears not to be the most important contributing factor to systematic error. In previous research where they applied a random sequence of PPT measurement in the lumbar region in healthy participants, the procedure with a hand-held manual analogue algometry and verbal command also resulted in significantly higher PPT score as compared to computerized algometry using a hand-held stop [34]. Further, both devices show excellent, almost perfect construct validity [11] compared to a force plate. So, the small but significant difference between the digital and analogue PPT values, but also the wide range between the lower and upper limit of agreement between both measurements, seems more likely to depend on method of measurement and the participant’s condition. Especially in people with migraine, we suspected that a wind-up phenomenon would be detectable due to consecutive PPT measurements, resulting in lower analogue PPT scores. Our results did not confirm this assumption. In contrast, we found higher analogue PPT values in both groups. The Bland-Altman plots revealed no proportional bias as with increased pressures, no larger (or smaller) differences were observed between the two devices.

The described variation in mean difference and limits of agreement of location between the groups can be explained by a larger variation in PPT scores in the group of people with migraine. Except for the temporalis muscle, all locations show larger mean differences in mean PPT scores in the migraine group. This difference in mechanosensitivity in people with migraine may explain the variation in mean differences and limits of agreement between the groups.

Only one assessor performed all measurements. This assessor was trained to perform the measurements and was selected from a pool of physiotherapy students enrolled in a master program. Although previous research showed high values for intratester and interrater reliability in novice and experienced assessors [35, 36], the lack of information on reliability of the measurement may hamper the interpretation of our results.

For interpretation of the ICC, we applied widely accepted cut-off points as described by Koo et al. [31]. According to their recommendation, we included the confidence interval around the ICC in the reported cut-off point. The concurrent validity of the digital and analogue algometer was excellent in healthy participants because ICC values ranged from 0.96 to 0.99 with the lower boundaries of the confidence intervals not below 0.90. For people with migraine, both algometers showed moderate to excellent ICC values ranging from 0.82 to 0.97 with the lowest boundary of the confidence interval not below 0.65.

Since the ICC 3.1 is not sensitive to systematic errors, the Bland-Altman plots and LoA are essential to judge if simultaneous or interchangeable use of both algometers is applicable in clinical practice and research. Based on our study, clinicians must be aware of systematic bias as visualized in the Bland-Altman plots and the wide range between the upper and lower limits of agreement in people with migraine.

To determine a difference in PPT measurement using both algometers, a difference of more than 37–55 kPa in healthy participants and a substantial difference of more than 81–130 kPa in people with migraine has to be detected, depending on the measured test location. Therefore, to minimize measurement error, we conclude that both algometers are not interchangeable to measure PPTs in people with migraine.

We have studied the concurrent validity of a digital (Type II, Somedic Electronics, Solna, Sweden) and analogue algometer (Force Dial FDK, Wagner Instruments, Greenwich, Connecticut). There are other digital algometers, such as the AlgoMed (Medoc) and analogue algometers, available. Because the method of measurement may differ between devices, we have to be careful to generalize our findings to other brands of algometers.

Conclusion

The digital and analogue algometers show moderate to excellent concurrent validity in healthy participants and people with migraine. We recommend clinicians and researchers to use the same type of algometer in clinical practice and research, for both cross-sectional and longitudinal use. The analogue algometer can be considered as a valid, easy to handle, and low-cost instrument to assess mechanical sensitivity in healthy participants and people with migraine.

Availability of data and materials

The datasets used and/or analyzed during the current study are available from the corresponding author on reasonable request.

Abbreviations

ICC:

Intraclass correlation coefficient

PPT:

Pressure pain threshold

kPa:

Kilo Pascal

kg/cm2 :

Kilogram per square centimetre

ICHD – III:

International Classification of Headache Disorders III

References

  1. Castien RF, van der Wouden JC, De Hertogh W (2018) Pressure pain thresholds over the cranio-cervical region in headache: a systematic review and meta-analysis. J Headache Pain 19(1):9

    Article  Google Scholar 

  2. Bezov D, Ashina S, Jensen R, Bendtsen L (2010) Pain perception studies in tension-type headache. Headache 51(2):262–271. https://doi.org/10.1111/j.1526-4610.2010.01768.x

    Article  PubMed  Google Scholar 

  3. Andersen S, Petersen MW, Svendsen AS, Gazerani P (2015) Pressure pain thresholds assessed over temporalis, Masseter, and frontalis muscles in healthy individuals, patients with tension-type headache and those with migraine -a systematic review. Pain 156(8):1409–1423

    Article  Google Scholar 

  4. Szikszay TM, Hoenick S, Von Korn K, Meise R, Schwarz A, Starke W, Luedtke K (2019) Which examination tests detect differences in cervical musculoskeletal impairments in people with migraine? A systematic review and meta-analysis. Phys Ther 99(5):549–569. https://doi.org/10.1093/ptj/pzz007

    Article  PubMed  Google Scholar 

  5. Nahman-Averbuch H, Shefi T, Schneider VJ, Li D, Ding L, King CD, Coghill RC (2018) Quantitative sensory testing in patients with migraine: a systematic review and meta-analysis. Pain 159(7):1202–1223. https://doi.org/10.1097/j.pain.0000000000001231

    Article  PubMed  Google Scholar 

  6. Castien R, Blankenstein A, De Hertogh W (2015) Pressure pain and isometric strength of neck flexors are related in chronic tension-type headache. Pain Physician 18(2):E201-5

  7. La Touche R, Paris-Alemany A, Mannheimer JS, Angulo-Diaz-Parreno S, Bishop MD, Lopez-Valverde-Centeno A, Fernandez-Carnero J (2013) Does mobilization of the upper cervical spine affect pain sensitivity and autonomic nervous system function in patients with cervico-craniofacial pain? A randomized-controlled trial. Clin J Pain 29(3):205–215. https://doi.org/10.1097/AJP.0b013e318250f3cd

    Article  PubMed  Google Scholar 

  8. Cuadrado ML, Aledo-Serrano A, Navarro P, Fernandez-Perez C, Gonzalez-Suarez I, Galan L, Fernandez-De-Las-Penas C (2015) Short-term effects of greater occipital nerve blockades in chronic migraine: a double-blind. Randomized. Placebo-controlled clinical trial. Cephalalgia 37(9):864–872

    Article  Google Scholar 

  9. Bevilaqua-Grossi D, Goncalves MC, Carvalho GF, Florencio LL, Dach F, Speciali JG, Chaves TC (2016) Additional effects of a physical therapy protocol on headache frequency. Pressure pain threshold. And Improvement Perception in Patients With Migraine and Associated Neck Pain: A Randomized Controlled Trial. Arch Phys Med Rehabil 97(6):866–874. https://doi.org/10.1016/j.apmr.2015.12.006

    Article  PubMed  Google Scholar 

  10. Fischer AA (1990) Application of pressure algometry in manual medicine. J Man Med 5:145–150

    Google Scholar 

  11. Vaughan B, McLaughlin P, Gosling C (2007) Validity of an electronic pressure algometer. Int J Osteopath Med 10(1):24–28. https://doi.org/10.1016/j.ijosm.2006.12.003

    Article  Google Scholar 

  12. Waller R, Straker L, O’Sullivan P, Sterling M, Smith A (2015) Reliability of pressure pain threshold testing in healthy pain-free young adults. Scand J Pain 9(1):38–41. https://doi.org/10.1016/j.sjpain.2015.05.004

    Article  PubMed  Google Scholar 

  13. Park G, Kim CW, Park SB, Kim MJ, Jang SH (2011) Reliability and usefulness of the pressure pain threshold measurement in patients with myofascial pain. Ann Rehabil Med 35(3):412–417. https://doi.org/10.5535/arm.2011.35.3.412

    Article  PubMed  PubMed Central  Google Scholar 

  14. Antonaci F, Sand T, Lucas GA (1998) Pressure algometry in healthy subjects: Inter-examiner variability. Scand J Rehabil Med 1:3–8

    Article  Google Scholar 

  15. Potter L, McCarthy C, Oldham J (2006) Algometer reliability in measuring pain pressure threshold over normal spinal muscles to allow quantification of anti-nociceptive treatment effects. Int J Osteopath Med 9(4):113–119. https://doi.org/10.1016/j.ijosm.2006.11.002

    Article  Google Scholar 

  16. FernĂ¡ndez-de-las-Peñas C, Ge H-Y, Arendt-Nielsen L, Cuadrado ML, Pareja JA (2007) The local and referred pain from myofascial trigger points in the temporalis muscle contributes to pain profile in chronic tension-type headache. Clin J Pain 23(9):786–792. https://doi.org/10.1097/AJP.0b013e318153496a

    Article  PubMed  Google Scholar 

  17. Headache Classification Committee of the International Headache Society (2018) The international classification of headache disorders. 3rd edition. Cephalalgia 38(1):1–211. https://doi.org/10.1177/0333102417738202

    Article  Google Scholar 

  18. Hjermstad MJ, Fayers PM, Haugen DF, Caraceni A, Hanks GW, Loge JH (2011) Studies comparing numerical rating scales. Verbal rating scales. And visual analogue scales for assessment of pain intensity in adults: A systematic literature review. J Pain Symptom Manage 41(6):1073–1093. https://doi.org/10.1016/j.jpainsymman.2010.08.016

    Article  PubMed  Google Scholar 

  19. Yang M, Rendas-Baum R, Varon SF, Kosinski M (2010) Validation of the Headache Impact Test (HIT-6) across episodic and chronic migraine. Cephalalgia 31(3):357–367. https://doi.org/10.1177/0333102410379890

    Article  PubMed  Google Scholar 

  20. Kosinski. M, Bayliss MS, Bjorner JB, Ware JE, Garber WH, Batenhorst A, Tepper S (2003) A six-item short-form survey for measuring headache impact: The HIT-6. Qual Life Res 12(8):963–974

  21. Fernandez-de-las-Penas C, Madeleine P, Cuadrado ML, Ge HY, Arendt-Nielsen L, Pareja JA (2009) Pressure pain sensitivity mapping of the temporalis muscle revealed bilateral pressure hyperalgesia in patients with strictly unilateral migraine. Cephalalgia 29(6):670–676. https://doi.org/10.1111/j.1468-2982.2008.01831.x

    Article  CAS  PubMed  Google Scholar 

  22. Fernandez-de-las-Penas C, Madeleine P, Caminero AB, Cuadrado ML, Arendt-Nielsen L, Pareja JA (2010) Generalized neck-shoulder hyperalgesia in chronic tension-type headache and unilateral migraine assessed by pressure pain sensitivity topographical maps of the trapezius muscle. Cephalalgia 30(1):77–86. https://doi.org/10.1111/j.1468-2982.2009.01901.x

    Article  CAS  PubMed  Google Scholar 

  23. Fernandez-de-las-Penas C, Cuadrado ML, Arendt-Nielsen L, Pareja JA (2008) Side-to-side differences in pressure pain thresholds and pericranial muscle tenderness in strictly unilateral migraine. Eur J Neur 15(2):162–168

    Article  CAS  Google Scholar 

  24. Kinser AM, Sands WA, Stone MH (2009) Reliability and validity of a pressure algometer. J Strength Cond Res 23(1):312–314. https://doi.org/10.1519/JSC.0b013e31818f051c

    Article  PubMed  Google Scholar 

  25. Fischer A (1987) Pressure algometry over normal muscles: standard values, validity, and reproducibility of pressure threshold. Pain 30(1):115–126. https://doi.org/10.1016/0304-3959(87)90089-3

    Article  PubMed  Google Scholar 

  26. Chesterton LS, Sim J, Wright CC, Foster NE (2007) Interrater reliability of algometry in measuring pressure pain thresholds in healthy humans, using multiple raters. Clin J Pain 23(9):760–766. https://doi.org/10.1097/AJP.0b013e318154b6ae

    Article  PubMed  Google Scholar 

  27. Prinsen CA, Mokkink LB, Bouter LM (2018) COSMIN guideline for systematic reviews of patient-reported outcome measures. Qual Life Res 27(5):1147–1157. https://doi.org/10.1007/s11136-018-1798-3

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  28. Stenneberg MS, Busstra H, Eskes M, Trij E, Cattrysse E, Scholten-Peeters GGM, de Bie RA (2017) Concurrent validity and interrater reliability of a new smartphone application to assess 3D active cervical range of motion in patients with neck pain. Musculoskelet Sci Prac 34:59–65

    Article  Google Scholar 

  29. Quek J, Brauer SG, Treleaven J, Pua YH, Mentiplay B, Clark RA (2014) Validity and intra-rater reliability of an android phone application to measure cervical range-of-motion. J Neuroeng Rehabil 11:65

    Article  Google Scholar 

  30. Mcgraw KO, Wong SP (1996) Forming inferences about some Intraclass correlation coefficients. Psychol Meth 1(1):30–46. https://doi.org/10.1037/1082-989X.1.1.30

    Article  Google Scholar 

  31. Koo TK, Li MY (2016) A guideline of selecting and reporting Intraclass correlation coefficients for reliability research. J Chiropr Med 15(2):155–163. https://doi.org/10.1016/j.jcm.2016.02.012

    Article  PubMed  PubMed Central  Google Scholar 

  32. Altman DG, Bland JM (1983) Measurement in medicine: the analysis of method comparison studies. Statistician 32(3):307–317. https://doi.org/10.2307/2987937

    Article  Google Scholar 

  33. Scholten-Peeters GGM, Coppieters MW, Durge TS, Castien RF (2020) Fluctuations in local and widespread mechanical sensitivity throughout the migraine cycle: a prospective longitudinal study. J Headache Pain 21(1):16. https://doi.org/10.1186/s10194-020-1083-z

    Article  PubMed  PubMed Central  Google Scholar 

  34. Koo T, Guo J, Brown C (2013) Test-retest reliability, repeatability, and sensitivity of an automated deformation-controlled indentation on pressure pain threshold measurement. J Manip Physiol Ther 36(2):84–90. https://doi.org/10.1016/j.jmpt.2013.01.001

    Article  Google Scholar 

  35. Walton DM, Macdermid JC, Nielson W, Teasell RW, Chiasson M, Brown L (2011) Reliability, standard error, and minimum detectable change of clinical pressure pain threshold testing in people with and without acute neck pain. J Orthop Sports Phys Ther 41(9):644–650. https://doi.org/10.2519/jospt.2011.3666

    Article  PubMed  Google Scholar 

  36. Jayaseelan DJ, Cole KR, Courtney CA (2020) Hand-held dynamometer to measure pressure pain thresholds: a double-blinded reliability and validity study. Musculoskelet Sci Pract 30:102268

    Google Scholar 

Download references

Acknowledgements

We thank the MSG Science Netwerk Physiotherapy (https://www.msg-sciencenetwerk.nl/) for supporting this research and the patients for participating in this research project.

Funding

This work was supported by a grant from the MSG Science Network Physiotherapy (MSG2020).

Author information

Authors and Affiliations

Authors

Contributions

GGMSP, MWC and RFC participated in study conception and design. GGMSP, RFC and TSCD performed the data collection and data analyses. All authors interpreted the data, wrote the manuscript, read and approved the final manuscript.

Corresponding author

Correspondence to René F. Castien.

Ethics declarations

Ethics approval and consent to participate

The study was approved by the Medical Ethical Committee of the VU Medical Centre, Amsterdam, The Netherlands (METC-2015-551). Written informed consent was taken for participation.

Consent for publication

Not applicable.

Competing interests

The authors declare that they have no competing interests.

Additional information

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Castien, R.F., Coppieters, M.W., Durge, T.S.C. et al. High concurrent validity between digital and analogue algometers to measure pressure pain thresholds in healthy participants and people with migraine: a cross-sectional study. J Headache Pain 22, 69 (2021). https://doi.org/10.1186/s10194-021-01278-8

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s10194-021-01278-8

Keywords